Titrator

Piston burette with exchangeable head

Motor driven piston burette for exchangeable heads for dosing and titration

The TITRONIC® 500 is the ideal piston burette for manual titrations, accurate dosing applications as well as the preparation of solutions. When used with TitriSoft 3.3 and TitroLine® 7000, 7750 and 7800 it acts as an automatic dispenser or a titration burette.

General features
• High visibility, full color display that can be easily viewed from a distance and at extreme angles.
• Reagent data is securely stored in the intelligent and interchangeable modules
• Includes up to three USB, one LAN and two RS232 ports for expansion and connection of devices
• Export the results as PDF or CSV, also to networks.
• User management/password

Specific features for the TITRONIC ® 500
• Manual titration with the hand controller (mouse)
• Titration rate can be adjusted to optimize titration speed and accuracy.
• Programmable automatic calculations with individual units
• Dosing with adjustable dosing and filling rates for optimize speed and accuracy
• Sample/solution preparation
• Automatic weight recording when balance is connected


Scope of delivery:

Basic unit with magnetic stirrer, with 20 ml exchangeable head, with keyboard, hand controller, stand rod and titration clamp Z 305, power supply 100-240 V

  • Technical Specification

Technical Specification

Interfaces:  3 × USB (2 x USB-A and 1 × USB-B), 2 × RS-232-C and 1 x LAN
Interchangeable burettes  5, 10, 20 ml or 50 ml intelligent interchangeable burette units
Resolution (Steps):  20.000 steps
Dosing accuracy: according to DIN ISO 8655, part 3:
accuracy: < 0.15 %
precision: < 0.05 - 0.07 % (in dependence of the used burette size)
Display 3.5 inches - 1/4 VGA TFT graphic color display with 320 x 240 pixels
Housing material Polypropylene
Dimensions 15.3 x 45 x 29.6 cm (W x H x D), height with interchangeable head
Front keyboard Polyester coated
Material interchangeable burettes Valve: PTFE/ETFE;
Cylinder: borosilikate glass 3.3 (DURAN®)
hoses (FEP, blue) 
Weight 2.3 kg for basic device, 3.5 kg für complete device with interchangeable head (with empty reagent bottle) 
Ambient conditions Ambient temperature: + 10 ... + 40 °C for operation and storage Humidity according to EN 61 010, Part 1: max. relative humidity 80 % for temperatures up to 31 °C, linear decrease down to 50 % relative humidity at a temperature of 40 °C
Power supply: 100-240 V; 50/60 Hz, power input: 30 VA
CE-sign:  EMC compatibility according to the Council Directive: 2004/108/EG; applied harmonized standards: EN 61326-1:2006
Low-voltage directive according to the Council Directive 2006/95/EG Testing basis EN 61 010, Part 1
Magnetic stirrer TM 235, with Power supply TZ 1855, for vessels up to 500 ml, agitator speed infinitely adjustable from 500 - 2000 r/min, for the connection to TitroLine® 7XXX and TITRONIC® 50