Measuring input pH/mV |
pH/mV-input with elektrode socket according to DIN 19262. BNC via adapter. Reference electrode 1 x 4 mm socket |
Measuring range pH/accuracy |
- 3,0… 17,00 / 0,05 +/- 1 digit |
Measuring range mV/accuracy |
- 1900…1900/ 1,0 +/- 1 digit |
Measuring input temperature: |
Connector for Pt 1000 / NTC 30 KOhm (Sockets: 2 × 4 mm) |
Measuring range temperature |
- 30 …115 °C / 0,5 K +/- 1 digit |
Interfaces: |
2 × USB (1 USB-A and 1 × USB-B), 2 × RS-232-C |
Dosing units |
20 ml; 50ml also usable |
Resolution (steps): |
8000 steps |
Dosing accuracy: |
according to DIN ISO 8655, part 3:
accuracy: < 0.15 %
precision: < 0.05 % |
Display |
3.5 inches - 1/4 VGA TFT graphic color display with 320 x 240 pixels |
Housing material |
Polypropylene |
Dimensions |
135 x 310 x 205 mm (W x H x D), height with dosing unit, without stirrer |
Front keyboard |
Polyester |
Material dosing unit/valve |
Valve: PTFE/ETFE;
Cylinder: borosilikate glass 3.3 (DURAN®)
Hoses: (FEP, blue)
UV Cover: Ultem |
Weight |
2 kg |
Ambient conditions |
Ambient temperature: + 10 ... + 40 °C for operation and storage Humidity according to EN 61 010, Part 1: max. relative humidity 80 % for temperatures up to 31 °C, linear decrease down to 50 % relative humidity at a temperature of 40 °C |
Power supply: |
100-240 V; 50/60 Hz, power input: 30 VA |
CE-sign: |
EMC compatibility according to the Council Directive: 2004/108/EG; applied harmonized standards: EN 61326-1:2006
Low-voltage directive according to the Council Directive 2006/95/EG Testing basis EN 61 010, part 1 |
Magnetic stirrer |
TM 50, for glass vessels up to maximale 500 ml, speed can be adjusted without discrete steps 0 - 1000 min-1, connection to TITRONIC® 300 and TitroLine® 5000, including stand rod holder TZ 3885 |
pH-Electrode |
Glass shaft, 3 x ceramic junction, gel electrolyte, Silamid®-reference system, temperature sensor NTC 30 kOhm, sphere membrane, A-glass, 1 m fixed cable with DIN + 4 mm banana plug, length 120 mm, 12 mm Ø, -5...+80 °C, 0...14 pH |