General features
• Color display that can be easily viewed from a distance and at extreme angles.
• Export the results as PDF or CSV on USB sticks.
• Pre-installed standard methods

Specific features of the TITRONIC® 300
• Manual titration with the manual button ("mouse") incl. automatic calculation of results
• Automatic weight recording when a balance is connected
• Automatic dosing
• Adjustable dosing and filling rate for each method
• Adjustable filling between each dose step
• The "intelligent filling funtion" checks whether a dosing step can be carried out completely without filling. This reduces the number of faulty operations during serial dosing process.
• Number of methods: 5

Scope of delivery
Basic unit with magnetic stirrer, with stand rod, with mouse, with titration clamp, with ready to use with 50 ml dosing unit TZ 3160 with tubes, screw threads and titration tip, with external power supply 100-240 V.

  • Technical Specification

Technical Specification

Interfaces:  2 × USB (1 USB-A and 1 × USB-B), 2 × RS-232-C
Dosing units 50 ml; 20ml also usable 
Resolution (steps):  8000 steps
Dosing accuracy: according to DIN ISO 8655, part 3:
accuracy: < 0.15 %
precision: < 0.05 %
Display 3.5 inches - 1/4 VGA TFT graphic color display with 320 x 240 pixels
Housing material Polypropylene
Dimensions 135 x 310 x 205 mm (W x H x D), height with dosing unit, without stirrer
Front keyboard Polyester
Material dosing unit/valve Valve: PTFE/ETFE;
Cylinder: borosilikate glass 3.3 (DURAN®)
Hoses:  (FEP, blue)
UV Cover: Ultem 
Weight 2 kg
Ambient conditions Ambient temperature: + 10 ... + 40 °C for operation and storage Humidity according to EN 61 010, Part 1: max. relative humidity 80 % for temperatures up to 31 °C, linear decrease down to 50 % relative humidity at a temperature of 40 °C
Power supply: 100-240 V; 50/60 Hz, power input: 30 VA
CE-sign:  EMC compatibility according to the Council Directive: 2004/108/EG; applied harmonized standards: EN 61326-1:2006
Low-voltage directive according to the Council Directive 2006/95/EG Testing basis EN 61 010, part 1
Magnetic stirrer TM 50, for glass vessels up to maximale 500 ml, speed can be adjusted without discrete steps 0 - 1000 min-1, connection to TITRONIC® 300 and TitroLine® 5000, including stand rod holder TZ 3885